1 of 3

Bently Nevada 3300/35-12-01-01-00-00 Six-Channel Temperature RTD Input Monitor

Bently Nevada 3300/35-12-01-01-00-00 Six-Channel Temperature RTD Input Monitor

Only 2 item(s) left in stock
  • Manufacturer: Bently Nevada

  • Product No.: 3300/35-12-01-01-00-00

  • Country of origin:United States

  • Product Type: Temperature Monitors

  • Payment: T/T, Western Union

  • Weight: 2800g

  • Dimensions: 241 mm x 50 mm x 280 mm (H x W x D)

  • Shipping port: Xiamen

  • Warranty: 12 months

Quantity
View full details

Description

The Bently Nevada 3300/35-12-01-01-00-00 is a 3300 Series Six-Channel Temperature RTD Input Monitor engineered to provide continuous, on-line temperature monitoring for industrial machinery. This specific configuration accepts input signals from six 3-wire 100 ohm platinum RTD sensors across a full-scale range of 0 to 150 degC. It performs continuous signal conditioning, drives individual channel recorder outputs, and includes epoxy-sealed alarm relays for machine protection. Operating within the Bently Nevada 3300 system rack, the unit displays monitored temperatures on a 4-digit LCD screen and provides status indication via front-panel LEDs.

Features

  • Monitors six independent temperature channels using resistance temperature detector (RTD) inputs.
  • Pre-configured for 3-wire 100 ohm platinum RTD sensors with an alpha coefficient of 0.00392.
  • Factory calibrated for a 0 to 150 degC full-scale operating range.
  • Equipped with an epoxy-sealed dual alarm relay module for Alarm 1 and Alarm 2 contact outputs.
  • Provides individual proportional recorder outputs (+4 to +20 mA default) for each channel.
  • Features a front-panel 4-digit LCD display alongside OK, Alarm 1, Alarm 2, and Bypass status LEDs.
  • Supports internal plug-in jumpers for configuring alarm delays, reset logic, and recorder output types.

Applications

  • Continuous temperature monitoring of machinery bearings, windings, and stator temperatures.
  • On-line RTD signal conditioning and alarm protection in Bently Nevada 3300 rack systems.
  • Industrial monitoring applications requiring localized digital readout and dedicated relay outputs.

Ordering Information

Ordering Code Component Selected Option Technical Specification
Model 3300/35 Six-Channel Temperature RTD Input Monitor
A: Full-scale Range 12 0 to 150 degC
B: Transducer Input 01 100 ohm platinum (a = 0.00392) (3-wire)
C: Alarm Relay 01 Epoxy-sealed relays
D: Agency Approval 00 Not required
E: Safety Barriers 00 None

Technical Specifications

Parameter Specification
Manufacturer Bently Nevada
Model 3300/35-12-01-01-00-00
Product Type Six-Channel Temperature RTD Input Monitor
System Compatibility Bently Nevada 3300 System
Number of Channels 6 independent channels
Transducer Type 3-wire 100 ohm platinum RTD (a = 0.00392)
Full-Scale Range 0 to 150 degC
Alarm Relay Module Epoxy-sealed relays
Agency Approval Option Not required
Safety Barrier Option None
Power Consumption Nominal 4 watts
Display 4-digit 7-segment non-multiplexing LCD with decimal point
Display Accuracy (+25 degC) Within +/-1 degC (+/-1.8 degF) with tenths digit; +/-1.5 degC (+/-2.7 degF) without tenths digit
Display Resolution 0.1 or 1 (user-programmable)
LED Status Indicators OK (green), Alarm 1 (red), Alarm 2 (red), Bypass (red)
Recorder Output Signal User-programmable: +4 to +20 mA, 0 to -10 Vdc, or +1 to +5 Vdc
Recorder Accuracy (+4 to +20 mA) +/-1.8% of signal, +/-0.09 mA offset (at +25 degC)
Voltage Output Impedance 100 ohm (minimum load resistance 10 k ohm)
Current Output Compliance 0 to +12 Vdc across load (0 to 600 ohm load for +4 to +20 mA)
Alarm Setpoint Adjustment Digitally adjustable 0 to +500 degC (0 to +1000 degF), repeatability +/-1 degC (+/-1.8 degF)
Operating Temperature 0 degC to +65 degC (+32 degF to +150 degF)
Storage Temperature -40 degC to +85 degC (-40 degF to +185 degF)
Relative Humidity Up to 95%, non-condensing
Space Requirements Two rack positions (positions 3 through 12 in 3300 rack)
Weight 2.8 kg (6.2 lbs.)

Connections/Interfaces

Interface / Connector Function
RTD Input Connector Terminal connections for 3-wire RTD sensor inputs (Channels 1 to 6)
Recorder Output Connector Terminal outputs for individual channel recorder signals (CH1-CH6: +, -S, COM)
Alarm Contact Connector Relay output contacts for Alarm 1 and Alarm 2 (NC, ARM, NO)

Installation Guidelines

  • Mount the module into two vacant positions of a 3300 rack (positions 3 through 12; positions 1 and 2 are reserved for Power Supply and System Monitor).
  • Land 3-wire RTD sensor wiring directly onto the channel terminal connectors on the rear input assembly.
  • Configure internal plug-in jumpers behind the front panel to adjust settings such as Alarm Delays (1, 3, or 6 seconds), Latching/Nonlatching reset modes, and Recorder Output types.
  • Use the front-panel MODE, CHANNEL STEP, ALARM 1, and ALARM 2 pushbuttons to review monitored values and setpoint levels without interrupting monitoring operations.

Compliance and Certifications

  • CE EMC Directive (Certificate of Conformity: 158710)
  • CE Low Voltage Directive (Certificate of Conformity: 135300)

Product Documentation

Technical Datasheet (PDF) Complete specifications and technical drawings.

Technical Datasheet

Bently Nevada 3300/35-12-01-01-00-00 Six-Channel Temperature RTD Input Monitor

What is the Bently Nevada 3300/35-12-01-01-00-00 used for?

The Bently Nevada 3300/35-12-01-01-00-00 is a six-channel RTD temperature input monitor used to provide continuous, on-line machinery temperature monitoring and alarm protection within a Bently Nevada 3300 rack system.

What sensor input type and temperature range are configured on the 3300/35-12-01-01-00-00?

This model is configured for 3-wire 100 ohm platinum RTD sensors (alpha = 0.00392) with a full-scale range of 0 to 150 degC.

What relay option is included in the 3300/35-12-01-01-00-00?

The device includes option 01, which features an epoxy-sealed dual alarm relay module for Alarm 1 and Alarm 2 contact outputs.

Where is the temperature monitor installed in a 3300 system rack?

The monitor occupies two rack positions and can be installed in any available positions from position 3 through 12 (positions 1 and 2 are reserved for the power supply and system monitor).

What power supply input is required by the 3300/35 monitor?

The module receives power from the 3300 system rack and has a nominal power consumption of 4 watts.

Global Express Shipping

  • Standard Delivery: 4-6 Business Days via DHL, FedEx, and UPS.
  • Express Dispatch: Same-day dispatch for in-stock orders placed before 2:00 PM (GMT+8).
  • Worldwide Coverage: Serving over 150 countries, including rapid delivery to Saudi Arabia and UAE.

Returns & Warranty

  • 30-Day Guarantee: Returns accepted for in-stock products in original, factory-sealed packaging.
  • 12-Month Warranty: Every industrial component is backed by our professional technical warranty.

Orders are processed and delivered Monday-Friday (excluding public holidays).


For full eligibility, restocking fees, and international return details, please view our official Refund & Return Policy .

TECHNICAL SPECIFICATIONS

Country of origin
United States

Recently Viewed Products

Technical knowledge

Micro Drive Specialist EMS Bets on Custom Engineering to Broaden Market Appeal

UK micro-drive specialist EMS is pairing Faulhaber high-torque gearheads and compact motion controllers with in-house custom engineering—aiming beyond catalog sales into tailored mechatronics.

Bringing Object-Oriented Thinking to PLC Programming

Copy-paste ladder does not scale. Object-oriented PLC techniques—user-defined structures and instructions—cut duplication, propagate fixes once, and keep large multi-device programs readable.

Why Process Safety Management Programs Fail Even When OSHA Boxes Are Checked

OSHA boxes checked is not the same as a living PSM system. Refinery history shows programs fail when procedures, training, contractors, and investigations drift from how the plant actually runs.

What Is SCADA? Supervisory Control Without Replacing the PLC

SCADA sits above PLCs and RTUs: it acquires field data, alarms operators, and issues supervisory commands across plants and remote assets—without replacing local deterministic control.

What Is a VFD? Variable Frequency Drives That Actually Move the Needle

A VFD turns fixed-frequency mains into controlled voltage and frequency so AC motors match real load—cutting pump and fan energy while tightening torque, soft-start, and braking behavior.

What Is a PLC? The Controllers That Still Run the Plant Floor

PLCs still run the plant floor: scan cycles, I/O, and IEC 61131-3 languages decide whether machines stay deterministic. Here is how controllers work, where they win, and where they strain.